Campbell-scientific CPEC200 Closed-Path Eddy-Covariance System Bedienungsanleitung Seite 109

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 112
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 108
Appendix H. CPEC200 Pump Replacement
FIGURE H-8. Inlet and outlet tubing reconnected to pump
FIGURE H-9. Pump side with inlet and outlet tubing connected
2. Reattach the pump to metal box with two self-tapping #6 screws on
the back of the pump module electronics plate (FIGURE H-6).
Be careful not to pinch the fan wires under the pump and do not
overtighten screws.
3. Reattach the pump connector to the pump electronics (FIGURE H-5).
4. Place the pump assembly back into the foam on the shell cover. Make
sure the fan is in the hole in the end and the small tab on the metal box
is seated in the slit in the foam.
NOTE
H-5
Seitenansicht 108
1 2 ... 104 105 106 107 108 109 110 111 112

Kommentare zu diesen Handbüchern

Keine Kommentare