CS525 ISFET pH Probe
8
7. Troubleshooting
The most common causes for erroneous pH data include:
poor sensor connections to the datalogger
damaged cables
scratched chip
Problem:
Output signal is at its maximum value.
Possible reason:
Probe is not in fluid.
Chip is polluted.
Diaphragm is polluted.
Chip is scratched.
Suggestion:
Put probe in fluid.
Clean probe using soapy water method.
Probe can not be fixed if chip is scratched.
Problem:
Probe response is very slow.
Possible reason:
Diaphragm is chipped or polluted
Suggestion:
Clean probe with soapy water method.
Problem:
Probe signal is drifting.
Possible reason:
Diaphragm can be dried out.
Chip is scratched.
Suggestion:
Soak probe for 10 minutes in saturated KC1 solution (CSI PN 16349).
Probe can not be fixed.