Seite 1 - NR-LITE2 Net Radiometer

NR-LITE2 Net RadiometerRevision: 3/13 Copyright © 1998-2013Campbell Scientific, Inc.

Seite 2

NR-LITE2 Net Radiometer Support Arm 26120 Pole Support Arm 26120Crossarm FIGURE 4-2. Mounting the NR-LITE2 onto a pole (top) and crossarm (bottom) v

Seite 3 - Warranty

NR-LITE2 Net Radiometer The SCWin example provided here is for no wind speed correction. SCWin also supports dynamic wind speed correction; refer to

Seite 4 - Assistance

NR-LITE2 Net Radiometer 3. Select NR-LITE2 Net Radiometer (no wind speed correction), and select the right arrow (in center of screen) to add it to th

Seite 5 - Table of Contents

NR-LITE2 Net Radiometer 5. Choose the outputs and then select Finish. 6. In the Save As window, enter an appropriate file name and select Save. 7

Seite 6 - Figures

NR-LITE2 Net Radiometer 5. Overview The NR-LITE2 is used for measuring solar and far infrared radiation balance. This balance is known as the net (t

Seite 7

NR-LITE2 Net Radiometer FIGURE 5-1. Electrical circuit for the NR-LITE2 Net Radiometer 5.2 Spectral Properties The upwards facing sensor is calibra

Seite 8 - 4. Quickstart

NR-LITE2 Net Radiometer Error Angle of Incidence FIGURE 5-2. Cosine response of a typical net radiometer 5.4 Sensitivity to Wind Speed The calibrat

Seite 9 - 4.2 Mounting

NR-LITE2 Net Radiometer Differences between the NR-Lite and CNR1 as a function of wind speed for daytime conditions during 7 May ─ 6 June, 1998. (a) N

Seite 10 - Diagram

NR-LITE2 Net Radiometer 6.1 Spectral Spectral range: 0.2 to 100 μm Detector type: Thermopile Detector protection: PTFE coating Detector profile: Con

Seite 11

NR-LITE2 Net Radiometer 7. Operation 7.1 Wiring The NR-LITE2 can be measured with a differential or single-ended channel on the datalogger; a differ

Seite 13

NR-LITE2 Net Radiometer TABLE 7-2. Datalogger Connections for Single-Ended Measurement Function Color CR10X, CR510, CR500 CR23X, 21

Seite 14 - 5. Overview

NR-LITE2 Net Radiometer the calibration factor into 1000. For example, if the calibration factor is 15.8 µV/(W m-2), the multiplier is: 1000/15.8 = 6

Seite 15 - 5.2 Spectral Properties

NR-LITE2 Net Radiometer 7.2.3.2 CR1000 Example Program with Wind Speed Correction Shown below is an example program written for the CR1000 datalogger

Seite 16 - Angle of Incidence

NR-LITE2 Net Radiometer VoltDiff(NR_Wm2,1,mv25,2,True,0,_60Hz,100,0) If WS_ms>=5 Then CNR_Wm2=NR_Wm2*(1+0.021286*(WS_ms-5)) Else CNR_Wm

Seite 17 - 6. Specifications

NR-LITE2 Net Radiometer 7.2.3.4 CR10X Example Program with Wind Speed Correction This example measures the NR-LITE2 using a single-ended input and in

Seite 18 - 6.3 Mechanical

NR-LITE2 Net Radiometer 4: Z=X+F (P34) 1: 1 X Loc [ Wspd_m_s ] 2: -5 F 3: 24 Z Loc [ scratch_1 ] 5: Z=X*F (P37) 1:

Seite 19 - 7. Operation

NR-LITE2 Net Radiometer 8. Maintenance The radiometer is an ‘all weather’ instrument and is very stable, but should be handled with care. It require

Seite 20 - 7.2 Datalogger Programming

NR-LITE2 Net Radiometer Conversely, if the sensor is warmer than your hand, the reading will decrease. 8. Check the radiometer’s sensitivity to therm

Seite 21 - 7.2.3 Example Programs

NR-LITE2 Net Radiometer 22

Seite 23 - ;in units of (W m

Warranty “PRODUCTS MANUFACTURED BY CAMPBELL SCIENTIFIC, INC. are warranted by Campbell Scientific, Inc. (“Campbell”) to be free from defects in materi

Seite 24

Campbell Scientific Companies Campbell Scientific, Inc. (CSI) 815 West 1800 North Logan, Utah 84321 UNITED STATES www.campbellsci.com • info@campbel

Seite 25

Assistance Products may not be returned without prior authorization. The following contact information is for US and international customers residing

Seite 26 - 9. Troubleshooting

Table of Contents PDF viewers: These page numbers refer to the printed version of this document. Use the PDF reader bookmarks tab for links to spec

Seite 27

Table of Contents ii 9. Troubleshooting...20 9.1 Checking Sensor Operation...

Seite 28

NR-LITE2 Net Radiometer 1. Introduction The NR-LITE2 is a high-output thermopile sensor which measures the algebraic sum of incoming and outgoing al

Seite 29

NR-LITE2 Net Radiometer 3.1 Ships With (1) WRR traceable calibration certificate (1) Mounting arm from original manufacturer (1) Bird stick from orig

Seite 30

NR-LITE2 Net Radiometer 4.2 Mounting The mounting bracket kit, pn 26120, is used to mount the NR-LITE2 directly to a vertical pipe, or to a CM202, CM