OBS500 Smart Turbidity Meter with ClearSensor™ Technology
universal response curve. However, there is residual nonlinearity that is
removed by calibration and by computation of a TU value with a 2nd-order
polynomial. This section explains how to do a turbidity calibration.
0 1000 2000 3000 4000
Tur bidity (TU)
0.0
0.2
0.4
0.6
0.8
1.0
Normalized OBS-3+ Response (OPV330 VCSEL)
FIGURE 7-5. Normalized response of OBS500 to AMCO Clear
®
turbidity. The inset shows the response function of a turbidity
sensor to high-sediment concentrations.
AMCO Clear
®
SDVB turbidity standards are used to calibrate an OBS500
sensor. SDVB standards are made for individual instruments. Standards made
for one model of turbidity meter cannot be used to calibrate a different model.