LP02 Pyranometer
The sensor should be mounted with the cable pointing towards the nearest
magnetic pole, e.g., in the Northern Hemisphere point the cable toward the
North Pole.
The CM225 Solar Sensor Mounting Stand is used to attach the LP02 to a
vertical pipe (1.0 – 2.1” OD) as shown in Figure 6.3-1. The LP02 includes a
base with three leveling screws, bubble level, and mounting screws.
Attach the LP02 to the CM225 as follows:
1. Loosely mount the pyranometer and fixture on the mounting arm, with the
leveling screws lightly touching the mounting plate. Do not fully tighten
the two mounting screws.
2. Turn the leveling screws as required to bring the bubble of the spirit level
within the ring. (For easy leveling first use the screw nearest the spirit
level.)
3. Tighten the mounting screws to secure the assembly in its final position.
Check that the pyranometer is still correctly leveled and adjust as
necessary.
LP02 Pyranometer
Bubble Level
(3) Leveling Screws
CM225 Solar
Sensor
Mounting Stand
(2) Mounting Screws
CM200 Series
Crossarm
FIGURE 6.3-1. LP02 Pyranometer Attached to CM225 Solar Sensor