PWS100Present Weather SensorRevision: 3/12Copyright © 2006-2012Campbell Scientific, Inc.

Seite 2

PWS100 Table of Contents vi C-3. PWS100 communication cable ... C-4 C-4. Enclosure wiring det

Seite 3 - Warranty

Section 8. Functional Description 8-14

Seite 4 - Assistance

Section 9. Maintenance 9.1 General The PWS100 Present Weather Sensor is a robust instrument that will provide years of uninterrupted weather monitor

Seite 5 - PWS100 Table of Contents

Section 9. Maintenance 9-2 It is advisable to use airduster to blow any loose dust and dirt from the lenses as a first step. Using a lint free lens c

Seite 6

Section 10. Troubleshooting 10.1 Introduction If your PWS100 seems to be operating incorrectly, there are a number of checks you can make to isolate

Seite 7

Section 10. Troubleshooting 10.2.3 Ice has formed in the end of the hoods The heaters may not be working. Enter the command mode of the sensor, run

Seite 8

Section 10. Troubleshooting checked that there are no spider’s webs on the instrument, especially in and around the detection volume and hoods of the

Seite 9 - List of Figures

Section 10. Troubleshooting 10-4

Seite 10 - List of Tables

Appendix A. PWS100 Output Codes The codes that are used in the PWS100 are taken from WMO Manual on Codes 1995 edition, Suppl. No. 5 (VIII.2005), Rec.

Seite 11 - Section 1. Introduction

Appendix A. PWS100 Output Codes Clouds generally dissolving or becoming less developed during the past hour 01 - - - - - State of sky on the whole

Seite 12

Appendix A. PWS100 Output Codes Blowing or drifting snow or sand, visibility less than 1 km 29 DRSA / BLSA / DRSN / BLSN - - - - FOG 30 FG - - - ≤1

Seite 13 - 2.2 Laser Safety

Section 1. Introduction The PWS100 is a laser-based sensor that measures precipitation and visibility by accurately determining the size and velocity

Seite 14

Appendix A. PWS100 Output Codes Freezing precipitation, slight or moderate2 47 -UP / UP P- / P P- / P i<5.34 - Freezing precipitation, slight

Seite 15

Appendix A. PWS100 Output Codes Rain and snow, light 67 -SNRA P- SR- i<1.84 - Drizzle and snow, light 67 -SNDZ P- SL- i<0.64 - Rain, drizz

Seite 16

Appendix A. PWS100 Output Codes Snow showers, moderate1 86 SHSN S S 1.0≤i<5.0 >0.5 mm Snow showers, heavy1 87 +SHSN S+ S+ i≥5.0 >0.5 mm

Seite 17 - Section 4. Overview

Appendix A. PWS100 Output Codes Subjective Observations (Trappes, Paris, France, 14-16 May 1997) and the Working Group on Surface Measurements (Genev

Seite 18

Appendix A. PWS100 Output Codes However it is also noted that non-mixed reports such as rain or snow may also contain a proportion of other types of

Seite 19 - Section 5. Specifications

Appendix A. PWS100 Output Codes TABLE A-6. Intensity bounds for rain, drizzle and snow. Rain Drizzle Snow Rain, Drizzle and Snow Light i<2.5

Seite 20 - 5.3 Optical Specifications

Appendix A. PWS100 Output Codes A-10

Seite 21

Appendix B. Wiring EARTH GROUND FIGURE B-1. Underside of DSP enclosure TABLE B-1. Cable identifier Connector Number Connector Description Connect

Seite 22

Appendix B. Wiring FIGURE B-2. DSP PCB to DSP enclosure connections B-2

Seite 23 - Section 6. Installation

Appendix C. Cable Selection C.1 Power Cable The PWS100 is provided pre-wired with a default 10 m power cable as shown in Figure C-1. This cable shou

Seite 24 - 6.2 Unloading and Unpacking

Section 1. Product Overview 1-2

Seite 25 - 6.3 Installation Procedures

Appendix C. Cable Selection WIRING INSTRUCTIONS COLOR DSP SCREEN G BLACK G RED 12V POWER +24 Vdc – GREEN A/C OR DC -24 Vdc – WHITE A/C OR D

Seite 26

Appendix C. Cable Selection C.2 Communication Cable The PWS100 has RS-232, RS-422 and RS-485 communications capability. Use a screened 2 × 0.22 mm2

Seite 27

Appendix C. Cable Selection WIRING INSTRUCTIONS DSP COLOR 9WAY TX-Z BLUE 2 CTS-Y YELLOW 8 RTS-B WHITE 7 RX-A GREEN 3 G BLACK 5 G485 RED N/C G SCR

Seite 28

Appendix D. Software Flowchart D-1

Seite 29 - 6.3.4 Basic Wiring

Appendix D. Software Flowchart D-2

Seite 30

Appendix D. Software Flowchart Filter FFT Channel 1Autocorrelation and DC removalFilter FFTChannel 2Autocorrelation and DC removalInverse FFT and sel

Seite 31 - 6.3.6 Communication Options

Appendix D. Software Flowchart Filter FFTChannel 1Autocorrelation and DC removalInverse FFT and select real componentFind correlation peak positionSe

Seite 32

Appendix D. Software Flowchart D-5

Seite 33

Appendix D. Software Flowchart Determine equivolumetric diameter and volumeDetermine size and velocity table co-ordinatesDetermine fuzzy logic values

Seite 34

Appendix E. Menu System Map E-1

Seite 35 - 6.3.9 Initial Settings

Section 2. Cautionary Statements 2.1 Sensor Unit Safety The PWS100 sensor has been checked for safety before leaving the factory and contains no int

Seite 36 - 6.4.3 Lightning Rod

Appendix E. Menu System Map E-2

Seite 37 - Section 7. Operation

Appendix E. Menu System Map WELCOME TO THE PWS100 SETUP MENUID {id}SN {sn}(0) message 0(1) message 1(2) message 2(3) set time & date(4) configura

Seite 38 - 7.3 Terminal Mode

Appendix E. Menu System Map E-4

Seite 40 - 7.4 PWS100 Menu System

Campbell Scientific Companies Campbell Scientific, Inc. (CSI) 815 West 1800 North Logan, Utah 84321 UNITED STATES www.campbellsci.com • info@campbel

Seite 41

Section 2. Cautionary Statements If the laser is operated outside of the housing then the following warning applies: INVISIBLE LASER RADIATION AVOID

Seite 42

Section 3. Initial Inspection Upon receipt of the PWS100, inspect the packaging and contents for damage. File damage claims with the shipping company

Seite 43

Section 3. Initial Inspection 3-2

Seite 44

Section 4. Overview The PWS100 Present Weather Sensor is a laser based sensor capable of determining precipitation and visibility parameters for auto

Seite 45

Section 4. Overview 4-2

Seite 46

Section 5. Specifications 5.1 Mechanical Specifications Measuring Area: 40 cm2 (6.2 in2) Housing Materials: Iridite NCP conversion coated alumin

Seite 48

Section 5. Specifications 5.3 Optical Specifications 5.3.1 Laser Head Specifications Laser Source: Near-infrared (IR) diode, eye safe Class 1M un

Seite 49

Section 5. Specifications 5.6.2 Precipitation Measurements Particle Size*: 0.1 mm to 30 mm (0.004 in to 1.18 in) Size Accuracy*: ± 5% (for parti

Seite 50

Section 5. Specifications 5-4 The particle buffer is able to hold raw data for 500 typical particles. The processor is able to process the particles

Seite 51

Section 6. Installation 6.1 Location and Orientation The PWS100 measures environmental variables and is designed to be located in harsh weather cond

Seite 52

Section 6. Installation In order to minimize user interaction with the unit, the PWS100 should be placed away from sources of contamination, in the c

Seite 53

Section 6. Installation 6.3 Installation Procedures 6.3.1 Assembling the PWS100 The PWS100 comes as a single unit, with the DSP enclosure attached

Seite 54

Section 6. Installation Bracket TabU-bolt Bracket DSP Plate FIGURE 6-2. Hardware for mounting the top of the DSP plate to a pole 6-4

Seite 55

Section 6. Installation Bracket Tab Notches FIGURE 6-3. Placing the PWS100 onto the bracket 6-5

Seite 56

Section 6. Installation FIGURE 6-4. PWS100 mounted to a mast or pole Ensure that the PWS100 is mounted according to Figures 6-2 through 6-4. Do not

Seite 57

Section 6. Installation 6.3.3 Connecting Cables The sensor unit comes with the DSP control unit fixed to the sensor arm. All cabling between the sen

Seite 58

Warranty “PRODUCTS MANUFACTURED BY CAMPBELL SCIENTIFIC, INC. are warranted by Campbell Scientific, Inc. (“Campbell”) to be free from defects in materi

Seite 59

Section 6. Installation PG9 CABLE GLAND PG11 CABLE GLAND(HOOD HEATER) LEMO 4-PIN (CONNECTOR FORCS215-PWS) EARTH GROUNDFIGURE 6-5. Underside of DS

Seite 60

Section 6. Installation 6.3.5 Desiccant The desiccant bags should be removed from the plastic bags in which they are shipped before placing them ins

Seite 61

Section 6. Installation FIGURE 6-7. Removal of DSP cover. FIGURE 6-8. Exposing the DSP board. 6-10

Seite 62

Section 6. Installation The location of the dip switches on the board is shown in Figure 6-9 and the dip switches themselves are shown in detail in F

Seite 63

Section 6. Installation FIGURE 6-9. DSP board dip switch location (circled) FIGURE 6-10. Dip switches (defaults set - 00011100) 6-12

Seite 64

Section 6. Installation 6.3.7 Installing Power Supply Power supply connections can be made in the PWS100 24 Vdc/12 Vdc power supply cabinet using th

Seite 65

Section 6. Installation 6-14 replacing the desiccant. This is of particular importance if using the sensor in corrosive or salt laden atmospheres. 6.

Seite 66

Section 7. Operation 7.1 Introduction The best way of becoming familiar with the sensor is to setup the sensor and connect it to a PC running Window

Seite 67

Section 7. Operation configuration options. These are discussed below within the context of setting up the sensor using a terminal emulator. 7.3 Te

Seite 68 - 0.178*D

Section 7. Operation In the descriptions which follow ↵ symbolizes the pressing of the ENTER key. Input parameters in italics should be user-defined

Seite 69

Assistance Products may not be returned without prior authorization. The following contact information is for US and international customers residing

Seite 70

Section 7. Operation The data storage function will not store data while the terminal or menu is active. This means data will be missing during these

Seite 71

Section 7. Operation Section 7.4.6 gives further details of how to set up the PWS100 with the command set in the terminal mode (option 7 from the SET

Seite 72

Section 7. Operation If option 1 is chosen from the MESSAGE menu then the system will display the MESSAGE PARAMETERS and FIELDS menu for that message

Seite 73

Section 7. Operation From the MESSAGE PARAMETERS and FIELDS menu if field 1 is chosen then the MESSAGE MODE menu (see Figure 7-5) will be displayed.

Seite 74

Section 7. Operation TABLE 7-2. Message Field parameters Message Field Parameter Output 34 External sensor (Aux) 40 Precipitation intensity (mmh-1

Seite 75 - Wetness grid value

Section 7. Operation Note that user defined messages cannot make field references to other user defined messages but can make reference to fixed mess

Seite 76

Section 7. Operation 7.4.1.3 Message Field 10 To 19 Fixed Messages No fixed messages have been defined yet. The user cannot change the fixed message

Seite 77

Section 7. Operation 10. DC voltage of upper detector greater than 1.5 V. This can be caused by sun directly shining into lens or a pws fault. 11.

Seite 78

Section 7. Operation The PWS100 only measures particles / visibility 90% of the time so precipitation intensity is scaled appropriately. NOTE 7.4.1.1

Seite 79 - 7.8 Maintenance Commands

Section 7. Operation 15 =>4.50 0.50 16 =>5.00 0.50 17 =>5.50 0.50 18 =>6.00 0.50 19 =>6.50 0.50 20 =>7.00 93.00 Particle speed

Seite 80

PWS100 Table of Contents PDF viewers: These page numbers refer to the printed version of this document. Use the PDF reader bookmarks tab for links t

Seite 81 - 7.9 Other Commands

Section 7. Operation 7 0.795 0.13 8 0.925 0.13 9 1.055 0.13 10 1.185 0.13 11 1.375 0.25 12 1.625 0.25 13 1.875 0.25 14 2.125 0.25 15 2.375 0.25 16 2.

Seite 82 - 7.10.1 Connections

Section 7. Operation 10 0.95 0.1 11 1.10 0.2 12 1.30 0.2 13 1.50 0.2 14 1.70 0.2 15 1.90 0.2 16 2.20 0.4 17 2.60 0.4 18 3.00 0.4 19 3.40 0.4 20 3.80

Seite 83

Section 7. Operation 8 =>0.70 0.1 9 =>0.80 0.1 10 =>0.90 0.1 11 =>1.00 0.2 12 =>1.20 0.2 13 =>1.40 0.2 14 =>1.60 0.2 15 =>1.8

Seite 84

Section 7. Operation 11 =>1.00 0.2 12 =>1.20 0.2 13 =>1.40 0.2 14 =>1.60 0.2 15 =>1.80 0.2 16 =>2.00 0.4 17 =>2.40 0.4 18 =>2

Seite 85

Section 7. Operation 7.4.1.24 Message Field 100 Upper, Lower LED temperature This field will output the sampled internal upper and lower LED tempera

Seite 86

Section 7. Operation 7.4.1.35 Message Field 154 Watchdog Count, Maximum Particles Per Second, Particles Not Processed, Time Lag This field will outp

Seite 87 - 8.2 Optical Measurement

Section 7. Operation FIGURE 7-6. Message field menu Choosing option 0 on the MESSAGE FIELD menu will display more output parameter options. Choosin

Seite 88 - Doublet lens Grating Laser

Section 7. Operation FIGURE 7-7. Delete message menu 7.4.2 Top Menu Option 3 (Set Time and Date) Choosing option 3 from the SETUP menu brings up t

Seite 89

Section 7. Operation 7.4.3 Top Menu Option 4 (Configuration) Choosing option 4 from the SETUP menu brings up the CONFIGURATION menu (see Figure 7-9)

Seite 90

Section 7. Operation FIGURE 7-10. PWS100 ID menu Option 2 of the configuration menu gives the TRH PROBE TYPE menu (see Figure 7-11). Choose the cor

Seite 91 - 8.4 PWS100 Control Unit

PWS100 Table of Contents 6.4 Grounding and Lightning Protection ... 6-14 6.4.1 Equipment Grounding...

Seite 92 - 8.6 Algorithm Description

Section 7. Operation Option 3 of the configuration menu gives the WETNESS PROBE TYPE menu (see Figure 7-12). Choose the correct wetness probe that is

Seite 93 - Dr izzle

Section 7. Operation Option 5 of the configuration menu gives the HOOD HEATER TEMPERATURE menu (see Figure 7-14). The value that is to be set needs t

Seite 94

Section 7. Operation Option 7 of the configuration menu gives the OUTPUT MODE menu (Figure 7-16). This refers to the use of packetized output. The me

Seite 95

Section 7. Operation FIGURE 7-17. Calibration warning screen Once confirm is typed followed by the return key, the CALIBRATION top menu (see Figure

Seite 96 - 8.6.4 Present Weather

Section 7. Operation FIGURE 7-19. Calibration disc constants menu Selecting option 1 on the CALIBRATION top menu will bring up the VIEW / ADJUST CA

Seite 97 - 8.6.5 Visibility

Section 7. Operation FIGURE 7-20. View / adjust calibration menu Option 9 of the configuration menu gives the TERMINAL MODE menu (see Figure 7-21).

Seite 98 - 8.7 Applications

Section 7. Operation Option 10 of the configuration menu gives the PSU SHUT DOWN VOLTAGE menu (Figure 7-22). Enter the PSU input voltage level below

Seite 99 - 8.8 Internal Monitoring

Section 7. Operation 7.4.5 Top Menu Option 6 (Weather and Alarm Parameters) Choosing option 6 from the SETUP menu brings up the WEATHER AND ALARM PA

Seite 100

Section 7. Operation The snow water content scaling factor (SWCF) can be adjusted by choosing option 4 on the WEATHER PARAMETERS menu. This will brin

Seite 101 - Section 9. Maintenance

Section 7. Operation FIGURE 7-26. Snow water content adjustment The mixed precipitation threshold value can be adjusted by choosing options 5-8 on

Seite 102 - 9.3 Calibration

PWS100 Table of Contents 7.4.1.29 Message Field 105 DSP PSU Voltage, Hood and Dew Heater % Duty ...

Seite 103 - Section 10. Troubleshooting

Section 7. Operation FIGURE 7-27. Mixed precipitation threshold adjustment 7.4.6 Top Menu Option 7 (Terminal) Choosing option 7 from the SETUP men

Seite 104

Section 7. Operation of the operating system or a hardware reset. The SDI-12 values of temperature and relative humidity will be -999.00 if no such i

Seite 105

Section 7. Operation FIGURE 7-30. Done menu 7.5 Message Related Commands Messages can be set for a variety of output types, including fixed standa

Seite 106

Section 7. Operation The Message_Interval parameter must also be defined from 0 second to 32767 seconds which is the rate at which the system will di

Seite 107

Section 7. Operation To set manual polling mode the MSET command should be of the form: MSET Message_ID 0 0 Message_Fields ↵ e.g., MSET 0 0 0 105

Seite 108

Section 7. Operation Higher codes indicate other type of data (stored every 10 seconds) where values a, b, and c have different meanings as indicated

Seite 109

Section 7. Operation 7.5.3 Viewing Data Output on the Command Line A number of records (defined by the number n) of uncollected historic m data (mes

Seite 110

Section 7. Operation To set the adjustable visibility limit parameters, type: SETPARAM vislim1 vislim2 vislim3 snowwater mixthreshold↵ The system w

Seite 111

Section 7. Operation The PWS100 will then display the configuration parameters on screen in the following format: Date and Time PWS100 product identi

Seite 112

Section 7. Operation Currently three other sensors can be connected to the PWS100. The sensor configured is changed by parameters TRH_Sensor (Tempera

Seite 113

PWS100 Table of Contents 8.3 Additional Sensor Connections... 8-3 8.3.1 Using a CS215-PWS

Seite 114

Section 7. Operation Select the Xmodem 1k protocol and select the OS file using the ‘Browse’ button. Once selected press the ‘Send’ button. The Xmode

Seite 115

Section 7. Operation 7.8.3 Running the Calibration The detection volume is calibrated using the PWC100 calibrator. The calibration fixture plate whi

Seite 116

Section 7. Operation To set the time without the date being altered the following should be used: TIME hh:mm:ss↵ 7.9.2 Resetting the System The sys

Seite 117 - Appendix B. Wiring

Section 7. Operation relative gains or losses of the two clocks that the logger will over time either receive an extra transmission from the sensor o

Seite 118

Section 7. Operation 'Main Program BeginProg 'Open Serial Port to the PWS100 SerialOpen (Com1,115200,3,0,10000) 'Commands can b

Seite 119 - Appendix C. Cable Selection

Section 7. Operation 'Open Serial Port to the PWS100 SerialOpen (Com1,115200,3,0,10000) 'Commands can be included here to setup the s

Seite 120 - WIRING INSTRUCTIONS

Section 7. Operation 7-50

Seite 121 - C.2 Communication Cable

Section 8. Functional Description 8.1 General The PWS100 Present Weather Sensor is an optical sensor using the best of scatter meter and disdrometer

Seite 122

Section 8. Functional Description FIGURE 8-1. Laser unit Rod lens Cylindrical lens Doublet lens Grating Laser FIGURE 8-2. Laser unit showing l

Seite 123

Section 8. Functional Description FIGURE 8-3. Sensor unit Plano-convex lens Photodiode Filter FIGURE 8-4. Sensor unit showing light path extents

Seite 124

PWS100 Table of Contents List of Figures 4-1. PWS100... 4

Seite 125

Section 8. Functional Description 8.3.1 Using a CS215-PWS on the PWS100 The CS215-PWS temperature and RH probe should be connected directly to one o

Seite 126 - Polycrystalline processing

Section 8. Functional Description 8.4 PWS100 Control Unit The PWS100 control unit, shown in Figure 8-5 as a block diagram, is a custom designed DSP

Seite 127

Section 8. Functional Description 8.6 Algorithm Description 8.6.1 Detecting and Classifying Precipitation The PWS100 has a structured detection vol

Seite 128

Section 8. Functional Description Signal to Pedestal Ratio Analysis01002003004005006001 1.5 2 2.5 3 3.5 4 4.5 5Signal to Pedestal RatioParticle Count

Seite 129 - Appendix E. Menu System Map

Section 8. Functional Description size/velocity value of 0 for drizzle, 0.4 for rain, 0.3 for snow flakes, 0.4 for graupel, 0.1 for ice pellet etc. T

Seite 130

Section 8. Functional Description The PWS100 has exceptional sensitivity to the start and end of precipitation events. To avoid falsely reporting pr

Seite 131

Section 8. Functional Description 8.6.3 Precipitation Accumulation Precipitation accumulation is calculated in millimeters over a specified time per

Seite 132

Section 8. Functional Description 8.6.4.2 Visibility Types Also by using the WMO SYNOP code table (4680) a visibility type can be defined. These typ

Seite 133

Section 8. Functional Description possible to measure the immediate surroundings, selecting appropriate parameters, which can be related to the envir

Seite 134

Section 8. Functional Description 8.8 Internal Monitoring The PWS100 has a number of internal checks including temperature analysis inside each head