Campbell-scientific CR1000 Measurement and Control System Bedienungsanleitung Seite 157

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 588
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 156
Section 7. Installation
157
BeginProg
Multiplier = .05
Offset = 0
LoadFieldCal(true) 'Load the CAL File, if possible
Scan(100,mSec,0,0)
'Simulate measurement by exciting channel VX1/EX1
ExciteV(Vx1,mV,0)
'Make the calibrated measurement
VoltSE(SaltContent,1,mV2500,6,1,0,250,Multiplier,Offset)
'Perform a calibration if CalMode = 1
FieldCal(1,SaltContent,1,Multiplier,Offset,CalMode,KnownSalt,1,30)
'If there was a calibration, store it into a data table
CallTable(CalHist)
NextScan
EndProg
7.8.1.5.3 Zero Basis (Option 4)
Case: A non-vented piezometer (water depth pressure transducer) needs its offset
zeroed before deployment. Because piezometer temperature and barometric
pressure have strong influences on the pressure measurement, their offsets need to
be zeroed as well. The relationship between absolute pressure, gage pressure, and
piezometer temperature is summarized in the linear equation,
pressure=G*(R
0
R
1
)+K*(T
1
T
0
)+(S
0
S
1
)
where,
G=gagefactor(0.036PSI/digitistypical)
R
0
=outputatthezerostate(outofwater)
R
1
=measurement
K=temperaturecorrectioncoefficient(0.04PSI/istypical)
T
0
=rtemperatureatthezerostate
T
1
=temperaturemeasurement
S
0
=barometricpressureatthezerostate
S
1
=barometricpressuremeasurement.
The following procedure determines zero offset of the pressure transducer, water
temperature, and barometric pressure readings. Use the external keyboard /
display or support software numeric monitor to change variable values as directed.
Seitenansicht 156
1 2 ... 152 153 154 155 156 157 158 159 160 161 162 ... 587 588

Kommentare zu diesen Handbüchern

Keine Kommentare