Section 8. Operation
342
Table 79. File-Control Functions
File-Control Functions Accessed Through
1
Datalogger support software (p. 77) Program Send command
2
Datalogger support software File Control (p. 454) utility
3
Device Configuration Utility (DevConfig) (p. 92) software
4
Manual with CF card (CRD: drive) or Campbell Scientific mass-storage media (USB: drive). See
Data Storage
(p. 332)
5
Automatic with CF card (CRD: drive) or Campbell Scientific mass-storage media (USB: drive)
and Powerup.ini. See Power-up
(p. 343)
6
CRBasic instructions (commands). See Data-Table Declarations (p. 475) and File Management
(p. 515)
and CRBasic Editor Help.
7
Datalogger support software Retrieve (p. 454) command
8.3.4.1 File Attributes
A feature of program files is the file attribute. Table CR1000 File Attributes (p. 342)
lists available file attributes, their functions, and when attributes are typically
used. For example, a program file sent via the support software Program Send
command, runs a) immediately ("run now"), and b) when power is cycled on the
CR1000 ("run on power-up'). This functionality is invoked because Program
Send
sets two CR1000 file attributes on the program file, i.e., Run Now and Run
on Power-up. When together, Run Now and Run on Power-up are tagged as
Run Always.
Note Activation of the run-on-power-up file can be prevented by holding down
the Del key on the external keyboard / display while the CR1000 is powering up.
Table 80. CR1000 File Attributes
Attribute Function Attribute for Programs Sent to CR1000 with:
Run Always
(run on power-up +
run now)
Runs now and on
power-up.
a) Send
(p. 454)
1
b) File Control
2
with Run Now & Run on Power-up
selected.
c) CF card (CRD: drive) or Campbell Scientific
mass-storage media (USB: drive) power-up
3
using
commands 1 & 13 (see table Powerup.ini Commands
(p. 345)
).
Run on Power-up
Runs only on
power-up
a) File Control
2
with Run on Power-up checked.
b) CF card (CRD: drive) or Campbell Scientific
mass-storage media (USB: drive) power-up
3
using
command 2 (see table Powerup.ini Commands
(p.
345)
).
Run Now
Runs only when
file sent to CR1000
a) File Control
2
with Run Now checked.
b) CF card (CRD: drive) or Campbell Scientific
mass-storage media (USB: drive) power-up
3
using
commands 6 & 14 (see TABLE. Powerup.ini
Commands
(p. 345) ). However, if CF card (CRD:
drive) or Campbell Scientific mass-storage media
(USB: drive) is left in, program loads again from CF
card (CRD: drive) or Campbell Scientific mass-
storage media (USB: drive).