EC155 CO
2
and H
2
O Closed-Path Gas Analyzer
Factory calibrated range
CO
2
: 0 to 1000 µmol·mol
-1
H
2
O: 0 mmol·mol
-1
to 37
o
C dewpoint
Analyzer temp: -30
o
to 50
o
C
Baro pressure: 70 to 106 kPa
CO
2
performance
Zero max drift
3
: ±0.55 mg·m
-3
·°C
-1
(±0.3 μmol·mol·°C
-1
)
Gain Drift: ±0.1% of reading·°C
-1
(maximum)
Sensitivity to H
2
O: ±5.6 x 10
-5
µmol CO
2
·mol
-1
H
2
O (max)
H
2
O performance
Zero max drift
3
: ±0.037 g·m
-3
·°C
-1
(±0.05 mmol·mol
-1
·°C
-1
)
Gain Drift: ±0.3% of reading·°C
-1
(maximum)
Sensitivity to CO
2
: ±0.05 mol H
2
O·mol
-1
CO
2
(maximum)
CSAT3A sonic measurement precision
4
u
x
: 1 mm·s
-1
u
y
: 1 mm·s
-1
u
z
: 0.5 mm·s
-1
Sonic temperature: 0.025
o
C
CSAT3A sonic accuracy
5
Offset error
u
x
, u
y
: < ±8 cm·s
-1
u
z
: < 4 cm·s
-1
Gain error
Wind vector ±5° horizontal: < ±2% of reading
Wind vector ±10° horizontal: < ±3% of reading
Wind vector ±20° horizontal: < ±6% of reading
CSAT3A sonic reporting range
Full scale wind: ±65.553 m/s
Sonic temperature: -50° to +60°C
Sample cell sensors
6
Barometer
Basic barometer
Accuracy
-30 to 0°C: ±3.7 kPa at -30°C, falling linearly to ±1.5
kPa at 0
o
C
0° to 50°C: ±1.5 kPa
Measurement rate: 10 Hz
Optional enhanced barometer:
Manufacturer: Vaisala
Model: PTB110
Accuracy
-30 to 0°C: ±0.15 kPa
Measurement rate: 1 Hz
Temperature sensor:
Manufacturer: BetaTherm
Model: 100K6A1A Thermistor
Accuracy: ±0.15
o
C (-30
o
to 50
o
C)
4